Table 1 - Troubleshooting Guide
SYMPTOM POSSIBLE CAUSE TREATMENT
Restricted airow/high inhalation resis-
tance through entire system
1. Cylinder valve not completely
opened.
1. Open valve. Check ll pressure
2. Cylinder valve requires service 2. Connect to a different cylinder
3. Conical lter (14) contaminated 3. Replace lter with new
4. Insufcient medium pressure 4. See below
Insufcient medium pressure
1. Low inlet pressure 1. Rell test cylinder
2. Main spring (7) weakened 2. Replace main spring
High intermediate pressure (leakage or
freeow through second stage)
1. HP seat (6) damaged or worn 1. Replace HP seat
2. Piston o-rings (3,5) damaged and
worn
2. Replace o-rings
3. Crown o-ring (11) damaged 3. Replace o-ring
4. Crown orice (12) sealing surface
damaged
4. Replace crown
5. Damaged piston (4) 5. Replace piston
Leakage of air from ambient ports of
piston cap
1. Piston shaft o-ring (5) damaged or
worn
1. Replace o-ring
2. Piston head o-ring (3) damaged or
worn
2. Replace o-ring
Leakage of air from inlet tting 1. Inlet tting o-ring (15) damaged 1. Replace o-ring
CAUTION: Recommended treatments which require disassembly of the regu-
lator must be performed during a complete overhaul, according to the pre-
scribed procedures for scheduled, annual service. Do not attempt to perform
partial service.
NOTE: This is a partial list of possible problems and recommended treatments. For
more information, contact Aqua Lung’s Technical Services Department for assis-
tance with problems not described here.